Community Pressure Temperature Humidity Calibration Laboratory (PTUCal)

Community Pressure Temperature Humidity Calibration Laboratory (PTUCal)

 

As of April 2026, this Lab can no longer be accessed through applications to AMOF. However, it remains available to the community through the University of Leeds. Contact Adam Leedham for further details.

 

The PTUCal Laboratory is situated in the AMOF Laboratories at the University of Leeds.

Pressure calibration is performed by comparison with a Paroscientific Model 745 portable pressure standard.

The temperature and humidity part of the laboratory is centered around a Michell Instrument HG-10 RELATIVE HUMIDITY CALIBRATION SYSTEM

HG-10 RELATIVE HUMIDITY CALIBRATION SYSTEM

  • The HG-10 is a completely automatic temperature and humidity calibration system, which can be used to calibrate humidity sensors over a wide range of temperatures and humidities.
  • PC control allows start, end values, ramp speeds and dwell times to be selected, following which the system automatically tracks the desired humidity/temperature profile and records the measured and reference values.
  • Dew-Point Generator – HG10
    • PC control interface, control & logging software, complete system integration,
      range 1 to 95%rh, temp. -10 to +50C (-40 to 90C dp)
  • Temperature Test Chamber
    • Temperature Test Chamber – 110 litre size (630hx400wx350mmd)
  • Dry Air Generation – PSD2
    • Pressure Swing Dryer, -80C dp nominal
  • S8000 Remote – High Performance Precision Dew Point Meter
    • A high precision cooled mirror dew point meter for the measurement of dew point temperatures over the range -40 to +90°Cdp.
    • The instrument offers the flexibility of a remote sensor that can provide accurate measurements in-situ in the area of interest or in an industrial process.
    • Measurement Performance
      • Range -40 to +90°C dp
      • Accuracy ± 0.1°C for dew point and temperature
      • Repeatability ± 0.05°C
      • Measurement units °C or °F dew point and temperature; %RH ; gm/m3; g/kg; ppmv; ppmw(SF6)

 

The full specifications:  HG10-97158-UK-Datasheet-V3

Being based in the AMOF laboratories in Leeds allows us to offer extensive bench space, power, and data infrastructure options.

The temperature and humidity test chamber has a working volume of

  • 550 x 550 x 320 mm
Arriving
  • On arrival, please go directly to the Fairbairn House’s main entrance. Use the intercom system to dial the extension number for the person you are visiting (phone numbers are listed in the glass panel to your right).
Location on Campus
  • The PTUCal Laboratory in the AMOF labs in Fairbairn House.
Getting to Campus
  • Parking at the university is very limited. There are several pay-and-display car parks available. If you need to park, it’s probably best to contact your host.

The instrument scientist must be made aware of all visits to the PTUCal Laboratory.

The user must attend an introduction to the PTUCal Laboratory briefing, given by instrument scientist.

The user must prepare their own risk assessment and work plan. A template is provided for this.

Emergency Provision
  • There should be at least one first aider on-site at all times
  • A phone is always available and there is always access to a vehicle.
  • First aid supplies are available on site.

Any user owned instruments needing delivery should be shipped to the following address

NCAS – PTUCal Laboratory
Fairbairn House
71-75 Clarendon Road
Leeds
LS2 9PH
UK

Mark all items for the attention of Adam Leedham.

Consumables
  • If bringing in additional instrumentation then all consumables and tools for that instrumentation should be provided.
Costs
  • Instrument Insurance
    • The user must have comprehensive insurance for any kit they bring
  • Public Liability Insurance
    • NCAS is not liable for any damage or injury arising from the deployment or operation of this instrument when unattended by the instrument scientists.
  • Power
    • For equipment that uses more than 3kW/hr, a charge will be made.
  • Gas Consumption
    • The PTUCal Laboratory has no gas handling capabilities.
  • Shipping
    • The user is liable for all costs arising from the shipping of any instruments both to and from the PTUCal Laboratory.